Datasheet: Flowline EchoPod DL14 Multi-Function Ultrasonic Level Transmitter

Code Description

 

Buy online for small quantities. For large quantities, projects and repeat business send a quotation request to: sales@birkettcontrols.com.au

 

The EchoPod DL14 is a general purpose ultrasonic multi-function level transmitter providing continuous level measurement up to 1.25m (49.2″) with a 4–20 mA analog signal output and four relays, configured via Flowline’s free WebCal software. Each relay can be configured for alarm, automatic fill or empty functions.

 

The embedded level controller can lower cost by replacing external control hardware, making it ideally suited for corrosive, sticky or dirty liquids across a wide range of small day tank, skid, IBC, sump and process tank level applications.

 

Both configurations feature a 1″ G (BSP) process connection and ship with a 1.2m (48″) cable. The DL14-11 bundle includes the Fob USB adapter required for sensor configuration via WebCal software — ideal where on-site configuration, re-ranging, or relay programming is needed. The DL14-10 is supplied without the Fob, suited to applications where configuration has already been completed or the Fob is already held on-site.

 

The key features are:

  • Level measurement, switch and control functions up to 1.25m (49″)
  • 4–20 mA analog signal output, two-wire (reversible: 4–20 mA or 20–4 mA)
  • Four programmable relays for switch, pump or valve control and fail-safety, configurable as:
    • 1 pump or valve with 3 alarms
    • 2 pumps lead-lag with 2 alarms
    • 2 pumps duplexing with 2 alarms
    • 4 independent switch point alarms
  • Narrow 50mm (2″) beam width and short 50mm (2″) dead band, optimised for small tanks
  • PVDF transducer and Type 6P polycarbonate enclosure for corrosive liquids
  • Automatic temperature compensation for accurate measurement
  • Fast and easy configuration via free WebCal PC software and USB Fob adapter
  • Configurable signal fail-safe: 4 mA, 20 mA, 21 mA, 22 mA or hold last
  • Relay contact fail-safe configurable: power loss hold last; power on open, close or hold last
  • CE, RoHS compliant; cFMus approved

 

Note: The Fob USB adapter (included with DL14-11) is required for initial configuration, relay programming or re-ranging via WebCal software. Fobs are universal and compatible with any WebCal-compatible Flowline sensor. If purchased separately, the Fob is Flowline P/N LI99-1001.

 

Delivery time for non-stocked configurations is typically 2-4 weeks.

 

To order a configuration not included in the above list or for application advice, send your request to: sales@birkettcontrols.com.au or contact us for more information on 0418 325 563

 

Typical Application Areas:

 

Level measurement, switching and control of liquids for:

  • Chemical feed, skid or machine, IBC or drum, and small waste sumps
  • Water and wastewater treatment
  • Municipal wastewater treatment — chemical solution dosing tanks with multi-stage alarm control
  • Cooling tower level control and monitoring
  • Pump lift stations and automated fill/empty applications
  • Food & beverage and pharmaceutical processing
  • Mining and industrial process tanks
Enclosure Material

Polycarbonate (Type 6P, encapsulated, corrosion resistant & submersible)

Transducer Material

PVDF

Process Connection

1" G (BSP)

Mount Gasket

Viton®

Enclosure Rating

Type 6P (IP68 equivalent)

Measuring Range

1.25m (49")

Dead Band

50mm (2")

Accuracy

±3mm (0.125")

Resolution

0.5mm (0.019")

Beam Width

50mm (2")

Process Temperature

–7°C to 60°C

Ambient Temperature

–35°C to 60°C

Max Working Pressure

2 bar (30 PSI)

Cable Length

1.2m (48"), 9-conductor shielded, polyurethane jacket

Strain Relief Material

Santoprene

Output Options

4–20 mA, two-wire (reversible); four SPST relays, 1A / 60 VA each

Relay Configuration Modes

Single set point alarm, or latched on two set points for automatic fill or empty in single or duplex control modes with fail-safe logic

Power Supply

24 VDC (loop powered), 0.5W consumption, 400Ω max loop resistance

Configuration

WebCal® PC software (free download), Windows®, USB 2.0 via Fob adapter (P/N LI99-1001); non-volatile memory

Compliance

CE, RoHS; cFMus